Checklist Rules

Your own checklist items, added on top of the Rimsys regulatory checklist. Active rules are applied automatically whenever a gap assessment runs on a matching submission.

5 rules
IDRuleApplies toSourceSeverityStatus
CR-001Subcontractor sterilisation validation must include EO residual test reports
EU MDR
Sterile, Class III
Section: Quality Management
Notified Body finding
DEKRA deficiency letter NB-2024-Q3-118 (12 Apr 2024)
BlockerActiveEdit
CR-002Quality agreement with each critical supplier must be cited by document number
All
All products
Section: Quality Management
Internal SOP
SOP-8812 Supplier Controls, rev. F
WarningActiveEdit
CR-003Declaration of Conformity must list the Basic UDI-DI
EU MDR
All products
Section: Declarations of Conformity
Notified Body finding
BSI technical review TR-2025-0441
BlockerActiveEdit
CR-004Accelerated aging protocol must state Q10 and chamber humidity
US FDA
Sterile, Vascular catheter
All sections
Authority deficiency
FDA AI request K241133 — Deficiency 4
BlockerActiveEdit
CR-005Cybersecurity SBOM must be attached for connected devices
US FDA
Software (SaMD)
All sections
Internal SOP
SOP-9020 Cybersecurity, rev. B
WarningDraftEdit

See it in action: open SUB-054 and run the AI Gap Assessment on section 1.4 — company rules appear in the Regulatory checklist with their own badge.